MEDICAL SUCTION UNIT

OB 1000 FA (BSU 220)

Max. Vacuum power

850 mbar (85 kPa, 638mmHg) ±10%

Vacuum regulation

linear

Vacuum range regulation

> 22 liters per minute with free air ±10%

Max. running time with the maximum current-load

Approximately 45 minutes ±10%

Approximate maximum noise energy

65 dBA

Accuracy of the analogue gauge

±5%

Accuracy of the battery power monitor

±5%

Running/Charging

12 ÷ 15 Vdc (direct current)

Max. current load

70W (max. current 6A)

Battery

sealed acid type, capacity 12V / 3,2 Ah

Max. time for charging

15 hours

Main safety

Internal not accessible to the user

Max. dimensions

320mm (width) x 100mm (depth) x 250mm (height)
12.6 in (width) x 3.94 in (depth) x 9.84 in (height)

Weight

3.5 Kg Max. complette with all accessories

Tolerance on all values

±5%

PRODUCT'S / OSCAR BOSCAROL - ITALY

MEDICAL SUCTION UNIT

OB 3000 FA (BSU 3000)

Max. Vacuum power

800 mbar (80 kPa, 600mmHg) ±10%

Vacuum regulation

linear with integrated mechanical regulator

Max. flow rate

33 LPM (liters per minute) at open flow ±10%

Vacuum range regulation

30 ÷ 800 mbar (3 ÷ 80 kPa)

Max. running time with the maximum current-load

60 minutes ±10% run time

Approximate maximum noise energy

70 dB

Accuracy of the analogue gauge

±5%

Accuracy of the battery power monitor

±5%

Running/Charging

11 ÷ 25 Vdc (direct current)

Max. current load

80W (max. current 8A)

Battery

Extractable, LiPo, 11.1V - 5Ah

Max. time for charging

10-15 hours consecutive

Main safety

Internal not accessible to operator

Max. dimensions

360mm (l) x 244mm (h) x 195mm (p)

Weight

2.6 Kg Max. complette with all accessories, 780gr wall bracket

Tolerance on all values

±5%

PRODUCT'S / OSCAR BOSCAROL - ITALY

MEDICAL SUCTION UNIT

OB MINI (BSU 360 UK)

Max. Vacuum power

850 mbar (85 kPa, 637.5mmHg) ±10%

Vacuum regulation

linear with integrated mechanical regulator

Max. flow rate

24 LPM (liters per minute) at open flow ±10%

Vacuum range regulation

30 ÷ 850 mbar (3 ÷ 85 kPa)

Max. running time with the maximum current-load

>70 minutes ±10% run time

Approximate maximum noise energy

70 dBA

Accuracy of the analogue gauge

±5%

Accuracy of the battery power monitor

±5%

Running/Charging

11 ÷ 25 Vdc (direct current)

Max. current load

80W (max. current 8A)

Battery

Extractable, LiPo, 11.1V - 5Ah

Max. time for charging

10-15 hours consecutive

Main safety

Internal not accessible to operator

Max. dimensions

290mm (l) x 240mm (h) x 95mm (d)

Weight

1.8 Kg Max. complette with all accessories

Tolerance on all values

±5%